HomeMy WebLinkAboutCERTIFICATE OF CALIBRATION% ertificate. of Callbratioh
# KELC-430603 IIIIIIIII�IIIulll�lulllllllllll� ��Il��uulll�� I��
Kimball Electronic Laboratory, Inc.
Precision Measurement Equipment Specialists
Calfbi"on'Perf6rmed
KIMBALL ELECTRONIC LABORATORY, INC
8081 W 21 LANE
HIALEAH, FL. 33016
Egwpmerit;lnfo_rmation KELt,f-D ,QUE77668 ^ i_
Description: OVEN v
Manufacturer: QUINCY LAB.
Model Number. 21-350
Part Number. 21-350
Range: 150-550 OF
Purchase Order #
QUEST ENGINEERING
2737 NW 19TH STREET
POMPANO BEACH FL 33069
Serial Number.. B23-3363
Customer I-D.: N/A
Cust. Barcode: N/A
Cust. Dept.: N/A
Specifications: MANUFACTURERS
This is to certify that the above listed instrument meets or exceeds a0 specifications as stated 1
been calibrated using measurement standards traceable to the National Institute of Standards
derived by the ratio type of self -calibration techniques. This calibration is in accordance with Ki
A21-A-Accredged to ISOIIEC-170252005 and compliantwith MILSTI) 45662A and ANSUNCS
uncertainty used to calculate the Test uncertainty Ratio, with a coverage factor of K=2 at a co
Cal Date: 17-Feb-17
Cal. Due Date: 17-Feb-18
Cal.lnterval: 12 MONTHS
Received: IN TOLERANCE
Calibration Result: PASS
Environmental Conditions: 75 DEG F / 45 % RH
Performed By: RAUL
GONZALEZ
Procedure: SYN1910
Cal Location: ON -SITE
n the referenced procedure at the points tested (unless otherwise notetl). I[ has
and Technology (NIST), or toNIST accepted Intrinsic standards of measurement or,
mbali Electronic Laboratory, Inc Quality Assurance Manual. KELI's Quality system is
L Z540.3-2006. TURS when applicable are greater than or equal to 4:1; with expanded
nfidence level of appmxfmately 95%. unless otherwise noted.
Any number of factors may cause the calibration item to drift out of calibration before the recommended interval has expired.
GabbratfonRemarks�:'� �� yam_; � , THIS UNIT WAS FOUND TO BE INTOLERANCE AT THE TIME OF CALIBRATION.
PERFORMED ROUTINE CAL. NO ADJUSTMENTS REQUIRED
Standards Used To CalLbrate Equiprent ,
Company I.D. Manufacturer ModeUPN Description
KIM001 1061 OMEGA CL23A CALIBRATOR THERMOMETER
SCANNED
BY
St. Lucie County
Serial Number Last Cal. Cal. Due Date
T-279508 07-Apr-I6 30-Apr-17
Certified by:
L
Quality Assurance:
f- •.-•^•^.-
RAUL
JAVIER
.lj .l _-.•
GONZALEZ
17-Feb-17 8:17:40 AN
SALCEIRO
20-Feb-17 8:18:24 AM
This report may not be reproduced, except in full, unless permission for the publication of an approved abstract is obtained in writing from KELI Labs., Inc
Kimball Electronic Laboratory, Inc. - 8081 W. 21st Lane - Hialeah, FL. 33016
Tel: 305-822-5792 - Toll Free: 800-393-1094 - Fax: 305-362-3125 - Web: www.kelilab&COrn
�ateof issue;. 20 Feb,17; .. Page of 2
Uertificate of Calibration
# KEL.C-430601
Kimball Electronic Laboratory, Inc.
Precision Measurement Equipment Specialists
KIMBALL ELECTRONIC LABORATORY, INC
8081 W 21 LANE
HIALEAH, FL. 33016
Equpment Information KE-0`1 D QUE47-666
Description: DIGITAL SCALE
Purchase Order#
For QUE739 _
QUEST ENGINEERING
2737 NW 19TH STREET
POMPANO BEACH FL 33069
Manufacturer. OHAUS Cal Date: 17-Feb-17
Model Number: ES30R Cal. Due Date: 17-Feb-18
Part Number: ES30R Cal. Interval: 12 MONTHS
Range: 0-30 KG Received: IN TOLERANCE
Calibration Result: PASS
Serial Number: 30010805 Environmental Conditions: 75 DEG F / 45 % RH
Customer I.D.: NIA Performed By: RAUL
Cust. Barcode: NIA GONZALEZ
Gust. Dept.: NIA Procedure: SYN913
Specifications: ± 0.01 KG Cal Location: ON -SITE
This is to certify that the above listed instrument meets or exceeds all specifications as stated In the referenced procedure at the points tested (unless otherwise noted). It has
been calibrated using measurement standards traceable to the National Institute of Standards and Technology (NIST), or to NIST accepted intrinsic standards of measurement, or
derived by the ratio type of selfcaGbre6on techniques. This calibration is in accordance with I0mba017echunic Laboratory, Inc Quality Assurance Manual. KELrs Quality system is
AZI.A-Accredited to I30/IEG770252005 and compliant with MILSTD-45662A and ANSUNCSL Z540.3-2006. TURS when applicable are greater than or equal to 4:1; with expanded
uncertainty used to calculate the Test Dncenainty Ratio, with a coverage factor of K=2 at a confidence level of approximately 95%, unless otherwise noted.
Any number of factors may muse the calibration item to drift out of calibration before the recommended Interval has expired.
CalibrattotRemarks� _ _ ,�w,
THIS UNIT WAS FOUND TO BE IN TOLERANCE AT THE TIME OF CALIBRATION.
PERFORMED ROUTINE CAL. NO ADJUSTMENTS REQUIRED
Standards Used To`Calbrate Egarpment,_- � � -,_
Company I.D. Manufacturer MOdelfPN Description
KIM001 7019 AMETEKT-150WEIGHTS
Serial Number
18830
Last Cal.
02-Jun-16
Cal. Due Date
30-Jun-17
Certified by: - Quality Assurance:
RAUL JAVIER y
GONZALEZ 17—Feb-17 8:14.51 AM BALCEIRO 20—Feb-17 8:15:43 AM
This report may not be reproduced, except in full, unless permission for the publication of an approved abstract is obtained in writing from KELI Labs., Inc.
Kimball Electronic Laboratory, Inc. - 8081 W. 21st Lane - Hialeah, FL. 33016
Tel: 305-822-5792 - Toll Free: 800-393-1094 - Fax: 305-362-3125 - Web: www.kelilabs.com
Date of issue:-;20 Feb47 r+Page 1 of 2
](LIJ�Ra! t
CERTIFICATE OF CALIBRATION
TRACEABLE TO THE UNITED STATES NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
AND PERFORMED IN ACCORDANCE WITH IS010012-2003
THIS DEVICE IS CERTIFIED TO BE WITHIN +/- 0.5% OF FULL SCALE READING AT EACH TEST POINT AT
THE TIME OF SHIPMENT
CERTIFICATE NUMBER: 24859
CALIBRATION DATE: 1/23/2018
PART AND CUSTOMER INFORMATION
MODEL #: DFG2W2000
SERIAL#: 20606019
j CUSTOMER: CLEARY INSPECTION SERVICE INC.
CUSTOMER#: CLE33460
CALIBRATION DATA
APPLIED EXPECTED
FULL
VALUE VALUE
RUN1
AVERAGE
SCALE
(LB) (LB) I
(LB)
(LB)
ERROR
' 0.00 0.00
0.000
0.000
0.000%
100.00
400.00 400.00
j i
402.000
402.000
0.100%
600.00
800.00 800.00
III! j
804.000
804.000
0.200%
1000.00
ji 1200.00 1200.00
1
1197.000
1197.000
-0.150%
I 1400.00
I I
1600.00 1600.00
1599.000
1599.000
-0.050%
1800.00
j
2000.00 2000.00
!
I
2004.000
2004.000
0.200%
CALIBRATION EQUIPMENT USED
SERIAL #1: 11706016
SERIAL #2:
j CAL EQUIP: COMTEN INDUSTRIES load cell CNR EA, SIN
! CNR EA 11706016 was calibrated at ComTen Industries
I
following to ASTM E-74. The uncertainty is 2
' CAL DATE: Lbf. Report No: 20300
.
j 6112/2017
I _ I
I= THIS EQUIPMENT IS CERTIFIED TRACEABLE TO THE UNITED STATES NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
i
CONDITIONS
TEMPERATURE 70•F
RELATIVE HUMIDITY: 31 %
iAS FOUND NOTES: (•Indicates that the partwas out of tolerance) I
VALIDATION r+�
VOI,�jJ
CALIBRATION CERTIFIED
BY: CC PERFORMED ,
Calibration not valid without CERTIFIED stamp and initials